In short: McIndoe dissecting forceps have fine tips and a guide pin, and both are easily damaged. Clean them promptly after use, inspect the tips, pin and spring after every reprocessing cycle, and store them so that the tips are protected. Take a pair out of service if the tips no longer meet evenly or the pin is bent.
For the pattern itself, see McIndoe dissecting forceps: features, versions and sizes. For all patterns, see our guide to dissecting forceps.
After use
- Remove visible soil soon after the procedure, before it dries.
- Keep fine forceps apart from heavy instruments so the tips are not knocked.
Cleaning
- Follow your facility’s reprocessing procedure and the instructions supplied with the instrument.
- Clean the serrations or teeth, the inner faces of the blades and the area around the guide pin, where soil collects.
- Use a soft brush. Wire brushes and abrasive pads damage the surface.
- Rinse and dry thoroughly before sterilisation.
Ultrasonic cleaning is a cleaning step. It does not replace sterilisation. For the full sequence, see surgical instrument cleaning procedures.
Inspection checklist
| Check | What to look for |
|---|---|
| Tips | They meet evenly along the gripping surface, with no gap and no overlap |
| Teeth, on toothed versions | They interlock without catching |
| Guide pin | It is straight and enters its hole freely as the forceps close |
| Blades | They are not bent or twisted |
| Spring | The blades open fully and close with even pressure |
| Surface | No staining, pitting or rust, especially at the spring end and around the pin |
Storage
- Use tip protectors or a tray with separate positions for fine instruments.
- Do not stack other instruments on top.
When to take a pair out of service
- The tips are misaligned or bent.
- The guide pin is bent, loose or missing.
- There is pitting or rust that cleaning does not remove.
For replacements, see the McIndoe listings in ordering McIndoe forceps.
Frequently asked questions
How do you check McIndoe forceps after cleaning?
Close them and check that the tips meet evenly, that the guide pin enters its hole freely, and that the surface is free of staining and pitting.
Can McIndoe forceps be cleaned in an ultrasonic cleaner?
Follow the instructions supplied with the instrument and your facility’s procedure. Ultrasonic cleaning is a cleaning step and does not replace sterilisation.
When should McIndoe forceps be replaced?
When the tips no longer meet evenly, the guide pin is damaged, or there is pitting or rust.